Further Development of a Robotic-Assisted Transfer Device.
Author | |
---|---|
Abstract | :
Background: The task of performing transfers, such as from a wheelchair to a bed, has a high risk of injury to both the caregiver and the person being transferred. Although mechanical transfer devices can reduce these risks, these devices are not meant for use in the community and they still place strain on the caregiver when used. Purpose: The aim of this study is to describe feedback gathered from focus groups of potential users of the Robotic-Assisted Transfer Device (RATD) and describe design changes aimed at preparing the device for the next step in the development process. Method: The RATD was transferred to a newer electric-powered wheelchair (EPW), key components were redesigned, and the control program was updated to increase the safety of the device. Two focus groups, one consisting of people with disabilities and the other consisting of clinicians and caregivers, were conducted to gather feedback from potential users. Results: Error checking, safety zones, a motor brake, and a new track helped increase the safety of the device. Sixty-three percent of the people with disabilities and 83% of caregivers surveyed said they would use the device. Conclusions: The results from the focus groups were positive and the design changes were successful, but more development is needed before the RATD can be marketed. |
Year of Publication | :
1969
|
Journal | :
Topics in spinal cord injury rehabilitation
|
Volume | :
23
|
Issue | :
2
|
Number of Pages | :
140-146
|
Date Published | :
1969
|
ISSN Number | :
1082-0744
|
DOI | :
10.1310/sci2302-140
|
Short Title | :
Top Spinal Cord Inj Rehabil
|